Endovascular stenting of an HVAD (TM) outflow graft pseudoaneurysm that exerts compression and kinking stenosis on the soft portion of the prosthesis

Elbayomi M, Weyand M, Uder M, May M, Steger K, Roth JP, Tandler R (2023)


Publication Type: Journal article

Publication year: 2023

Journal

Book Volume: 11

Journal Issue: 5

DOI: 10.1002/ccr3.7349

Abstract

Key Clinical MessageComplex presentations of MCS patients may necessitate a multidisciplinary approach involving HF cardiologists, CT surgeons, advanced cardiac imagers, and interventional cardiologists in order to define the optimal management strategy. Left ventricle assist devices (LVADs) provide life-sustaining treatment for patients with terminal heart failure, but their intricacy allows for complications. One complication is LVAD outflow graft obstruction due to the graft's intraluminal thrombus or extraluminal compression. It may be treated endovascularly with stenting. We report an endovascular stenting of an outflow tract in HVAD (TM) (HeartWare Inc.) due to a pseudoaneurysm causing compression and kinking stenosis.

Authors with CRIS profile

How to cite

APA:

Elbayomi, M., Weyand, M., Uder, M., May, M., Steger, K., Roth, J.-P., & Tandler, R. (2023). Endovascular stenting of an HVAD (TM) outflow graft pseudoaneurysm that exerts compression and kinking stenosis on the soft portion of the prosthesis. Clinical Case Reports, 11(5). https://doi.org/10.1002/ccr3.7349

MLA:

Elbayomi, Mohamed, et al. "Endovascular stenting of an HVAD (TM) outflow graft pseudoaneurysm that exerts compression and kinking stenosis on the soft portion of the prosthesis." Clinical Case Reports 11.5 (2023).