Portable measurement system for silicon nanowire field-effect transistor-based biosensors

Voitsekhivska T, Suthau E, Wolter KJ, Baraban L, Zörgiebel F, Cuniberti G (2013)


Publication Type: Conference contribution

Publication year: 2013

Pages Range: 349-354

Conference Proceedings Title: Proceedings of the International Spring Seminar on Electronics Technology

Event location: ROU

ISBN: 9781479900367

DOI: 10.1109/ISSE.2013.6648271

Abstract

The aim of this study is the development of a portable measurement system that simplifies the analysis of bio-sensitive silicon nanowire field effect transistors (SiNW - FETs). In order to obtain the pH - dependent transfer characteristics of the Schottky barrier (SB) FETs, Id = f (Vg), a measurement unit and two biochip adapters are proposed. For the parallel measurement of multiple SiNW - FETs, an integrated CMOS analog multiplexer has been designed, allowing to connect of up to 32 transistors individually. In order to determine the influence of the measurement structure on signal integrity, overall size and system cost, an additional discrete analog multiplexer will be compared with the integrated solution. © 2013 IEEE.

Involved external institutions

How to cite

APA:

Voitsekhivska, T., Suthau, E., Wolter, K.J., Baraban, L., Zörgiebel, F., & Cuniberti, G. (2013). Portable measurement system for silicon nanowire field-effect transistor-based biosensors. In Proceedings of the International Spring Seminar on Electronics Technology (pp. 349-354). ROU.

MLA:

Voitsekhivska, T., et al. "Portable measurement system for silicon nanowire field-effect transistor-based biosensors." Proceedings of the 2013 36th International Spring Seminar on Electronics Technology, ISSE 2013, ROU 2013. 349-354.

BibTeX: Download