Bon V, Senkovska I, Wallacher D, Heerwig A, Klein N, Zizak I, Feyerherm R, Dudzik E, Kaskel S (2014)
Publication Type: Journal article
Publication year: 2014
Book Volume: 188
Pages Range: 190-195
DOI: 10.1016/j.micromeso.2013.12.024
A new instrumentation for in situ X-ray powder diffraction allows to follow structural changes in the crystal structure of flexible coordination polymers during adsorption and desorption of gases. The system is based on closed cycle He-cryostat that ensures the high temperature stability in the range of 5.5-450 K. The automated gas dosing system facilitates the physisorption measurements in the pressure range of 0.01-130 kPa. The system is implemented on KMC-2 and MAGS beamlines at Helmholtz Zentrum Berlin für Materialien und Energie. Using new instrumentation, the adsorption of CO
APA:
Bon, V., Senkovska, I., Wallacher, D., Heerwig, A., Klein, N., Zizak, I.,... Kaskel, S. (2014). In situ monitoring of structural changes during the adsorption on flexible porous coordination polymers by X-ray powder diffraction: Instrumentation and experimental results. Microporous and Mesoporous Materials, 188, 190-195. https://doi.org/10.1016/j.micromeso.2013.12.024
MLA:
Bon, Volodymyr, et al. "In situ monitoring of structural changes during the adsorption on flexible porous coordination polymers by X-ray powder diffraction: Instrumentation and experimental results." Microporous and Mesoporous Materials 188 (2014): 190-195.
BibTeX: Download