Objectives To evaluate the feasibility, accuracy, usability and acceptability of two upper arm self-sampling devices for measurement of autoantibodies and C reactive protein (CRP) levels in patients with immune-mediated rheumatic diseases (IMRDs). Methods 70 consecutive patients with IMRD with previously documented autoantibodies were assigned to supervised and unsupervised self-collection of capillary blood with the Tasso+ or TAP II device. Interchangeability of 17 biomarkers with standard venesection was assessed by: concordance, correlation, paired sample hypothesis testing and Bland-Altman plots. Patients completed an evaluation questionnaire, including the System Usability Scale (SUS) and Net Promoter Score (NPS). Results While 80.0% and 77.0% were able to safely and successfully collect capillary blood using the Tasso+ and TAP II within the first attempt, 69 of 70 (98.6%) patients were successful in collecting capillary blood within two attempts. Concordance between venous and capillary samples was high; 94.7% and 99.5% for positive and negative samples, respectively. For connective tissue disease screen, anti-Ro52 and anti-proteinase 3 autoantibody levels, no significant differences were observed. Self-sampling was less painful than standard venesection for the majority of patients (Tasso+: 71%; TAP II: 63%). Both devices were well accepted (NPS; both: +28%), usability was perceived as excellent (SUS; Tasso+: 88.6 of 100; TAP II: 86.0 of 100) and 48.6 %/62.9% of patients would prefer to use the Tasso+/TAP II, respectively, instead of a traditional venous blood collection. Conclusions Remote self-collection of capillary blood using upper arm-based devices for autoantibody and CRP analysis in patients with autoimmune rheumatic diseases is feasible, accurate and well accepted among patients.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Zarbl, J., Eimer, E., Gigg, C., Bendzuck, G., Korinth, M., Elling-Audersch, C.,... Knitza, J. (2022). Remote self-collection of capillary blood using upper arm devices for autoantibody analysis in patients with immune-mediated inflammatory rheumatic diseases. RMD Open, 8(2). https://doi.org/10.1136/rmdopen-2022-002641

MLA:

Zarbl, Joshua, et al. "Remote self-collection of capillary blood using upper arm devices for autoantibody analysis in patients with immune-mediated inflammatory rheumatic diseases." RMD Open 8.2 (2022).

BibTeX: Download