Reliability and validity of 4D rasterstereography under dynamic conditions

Betsch M, Wild M, Jungbluth P, Hakimi M, Windolf J, Haex B, Horstmann T, Rapp W (2011)


Publication Type: Journal article

Publication year: 2011

Journal

Book Volume: 41

Pages Range: 308-312

Journal Issue: 6

DOI: 10.1016/j.compbiomed.2011.03.008

Abstract

Purpose of this study was to evaluate the reliability and validity of 4D rasterstereography under dynamic conditions. Therefore simulated anatomical fixed points on a wooden plate were measured during different movements. Seven different motion patterns in all three angles of space were evaluated. The simulated parameter trunk length was measured with an accuracy of 3.58. mm (SD±3.29. mm) and the dimple distance was detected with an accuracy of 0.88. mm (SD±1.04. mm). With this rasterstereographic prototype it is possible to examine dynamically the spinal posture with adequate accuracy. © 2011 Elsevier Ltd.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Betsch, M., Wild, M., Jungbluth, P., Hakimi, M., Windolf, J., Haex, B.,... Rapp, W. (2011). Reliability and validity of 4D rasterstereography under dynamic conditions. Computers in Biology and Medicine, 41(6), 308-312. https://dx.doi.org/10.1016/j.compbiomed.2011.03.008

MLA:

Betsch, Marcel, et al. "Reliability and validity of 4D rasterstereography under dynamic conditions." Computers in Biology and Medicine 41.6 (2011): 308-312.

BibTeX: Download