AAPM Task Group Report 290: Respiratory motion management for particle therapy

Li H, Dong L, Bert C, Chang J, Flampouri S, Jee KW, Lin L, Moyers M, Mori S, Rottmann J, Tryggestad E, Vedam S (2022)


Publication Type: Journal article

Publication year: 2022

Journal

DOI: 10.1002/mp.15470

Abstract

Dose uncertainty induced by respiratory motion remains a major concern for treating thoracic and abdominal lesions using particle beams. This Task Group report reviews the impact of tumor motion and dosimetric considerations in particle radiotherapy, current motion-management techniques, and limitations for different particle-beam delivery modes (i.e., passive scattering, uniform scanning, and pencil-beam scanning). Furthermore, the report provides guidance and risk analysis for quality assurance of the motion-management procedures to ensure consistency and accuracy, and discusses future development and emerging motion-management strategies. This report supplements previously published AAPM report TG76, and considers aspects of motion management that are crucial to the accurate and safe delivery of particle-beam therapy. To that end, this report produces general recommendations for commissioning and facility-specific dosimetric characterization, motion assessment, treatment planning, active and passive motion-management techniques, image guidance and related decision-making, monitoring throughout therapy, and recommendations for vendors. Key among these recommendations are that: (1) facilities should perform thorough planning studies (using retrospective data) and develop standard operating procedures that address all aspects of therapy for any treatment site involving respiratory motion; (2) a risk-based methodology should be adopted for quality management and ongoing process improvement.

Authors with CRIS profile

Involved external institutions

How to cite

APA:

Li, H., Dong, L., Bert, C., Chang, J., Flampouri, S., Jee, K.-W.,... Vedam, S. (2022). AAPM Task Group Report 290: Respiratory motion management for particle therapy. Medical Physics. https://doi.org/10.1002/mp.15470

MLA:

Li, Heng, et al. "AAPM Task Group Report 290: Respiratory motion management for particle therapy." Medical Physics (2022).

BibTeX: Download