Zagrada E, Krickau T, Rascher W (2019)
Publication Type: Journal article
Publication year: 2019
Book Volume: 231
Pages Range: 67-73
Journal Issue: 2
DOI: 10.1055/a-0747-6253
Background As an alternative to ambulatory 24-h blood pressure monitors, the continuous, non-invasive blood pressure recording using pulse transit time (PTT) was compared with both oscillometric and intra-arterial measurement. Methods In 144 pediatric patients (9.4 +/- 5.2 years) acute blood pressure was determined using three different methods of measurement. In 57 patients (11.4 +/- 4.9 years) of the pediatric normal ward, the blood pressure was simultaneously determined continuously by means of PTT compared to the intermittent oscillometric long-term measurement. In 9 patients (9.8 +/- 6.8 years) of the pediatric intensive care unit with continuous intra-arterial blood pressure measurement, the blood pressure was measured in parallel by means of PTT.
APA:
Zagrada, E., Krickau, T., & Rascher, W. (2019). Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents. Klinische Pädiatrie, 231(2), 67-73. https://doi.org/10.1055/a-0747-6253
MLA:
Zagrada, Elena, Tobias Krickau, and Wolfgang Rascher. "Continuous Non-Invasive Blood Pressure Measurement Using Pulse Transit Time in Children and Adolescents." Klinische Pädiatrie 231.2 (2019): 67-73.
BibTeX: Download