Field simulation of applicators for interstitial microwave hyperthermia

Schaller G, Erb J, Engelbrecht R, Seegenschmiedt H (1995)


Publication Language: English

Publication Type: Conference contribution, Conference Contribution

Publication year: 1995

Publisher: IEEE

DOI: 10.1109/mwsym.1995.406045

Abstract

The electromagnetic field near clinically applied coaxial microwave applicators for interstitial hyperthermia is calculated using a finite integration algorithm program. Simulations of E-field and SAR distributions of different conventional applicators in a muscle phantom are presented and compared with measurements. The applied calculation method allows also design optimization and improvement of more sophisticated applicator types, e.g. those using a triaxial technique

Authors with CRIS profile

How to cite

APA:

Schaller, G., Erb, J., Engelbrecht, R., & Seegenschmiedt, H. (1995). Field simulation of applicators for interstitial microwave hyperthermia. In Proceedings of the IEEE MTT-S International Microwave Symposium. IEEE.

MLA:

Schaller, G, et al. "Field simulation of applicators for interstitial microwave hyperthermia." Proceedings of the IEEE MTT-S International Microwave Symposium IEEE, 1995.

BibTeX: Download